FAST: Hardware updates
Created 09/03/03 by EF
Updated 01/26/12 by EF
FAST CCD
John Geary and Steve Amato mounted and tested a
new CCD (FAST3, a UA
STA520A 2688x512 chip) in a dewar in Cambridge. The CCD and dewar
arrived on the mountain on 14 October 2005. We found problems with
the chip that required the dewar to travel back to Cambridge. As of
April 2006, the chip showed unexplained behavior that resulted in
lower sensitivity than for FAST2. Tests showed the problem was
in the software. Ted debugged the CCD controls, with the happy
result that the CCD is in good shape as of June 2006.
As of the end of June 2006, FAST3 (S/N 4377), has replaced FAST2. It requires
no UV-flooding, a significant advantage for Ridge operations.
FAST optics
Warren Brown had replacement optics for FAST re-coated during July 2003.
He replaced and aligned the internal optics during August
2003 shutdown. He also replaced a sol-gel recoated corrector. The
result was a marked increase in throughput (see the plot
here), especially at blue wavelengths. Throughput measurements were
made with standard stars and a 5" slit. The increase in blue throughput
has revealed that we have significant second-order contamination in FAST
spectra. The plot shown
here shows the
spectrum of the blue standard
BD+28 without the WG360 order-blocking filter (3700 A)
divided by the spectrum through the filter (thanks to Tom Matheson).
If you are concerned with spectrophotometry, you may wish to use
the WG360 filter for your observations.
We periodically observe standards with the 5" aperture to monitor the
throughput of the telescope and FAST. See the plots
here and
here for Warren's most recent
analysis, with FAST3 starting July 2006.
Re-aluminizing the primary in August 2005
returned the total system
throughput close to what we measured in September 2003, immediately after the
FAST refurbishment. Dusting off the internal optics of FAST (carefully
blowing and vacuuming) in early November 2005
yielded a nearly 10% throughput improvement. By April 2006, we were back
to the levels preceding the re-aluminization in 2005. Dusting off the
internal optics of FAST is now on a monthly schedule, as for the
corrector.
FAST alignment and corrector cleaning procedures
08/22/2003 Warren Brown
Equipment for FAST Alignment:
Optical bench, mounting blocks, tie-down bars, long threaded rods.
Laser, 24" post, post holder, tip/tilt stage, translation stage.
FAST Power supply, control box, laptop with serial interface and FAST.BAS.
Allen set, crescent wrench, needlenose pliers, small phillips and standard
screwdrivers, tape, scissors.
FAST Alignment procedure:
- Move FAST to optical bench.
The optical bench in August 2003 was located in the old Gamma-ray
control shed, now the HAT shed. I called Karen to get a forklift
operator. We used the forklift to pick up FAST in its rack and carry it
down to the shed. We then used the forklift to pick up FAST from its
aluminum handles, lift it out of the rack, and set the spectrograph on a
piece of wood + foam on the doorstep of the shed. After removing the
handles, we tipped the flange through the doorway and onto the optical
bench. Then we lifted and carried the opposite end up onto the bench.
- Mount FAST on optical bench.
Use the 3 aluminum mounting blocks as shown. The two short ones
go near the flange, with the 1/4-20 bolt holes facing away from the
instrument. The tall aluminum block is placed on the far end. The blocks
should only touch the metal rim, not the sidewalls, of the spectrograph.
Finally, use the long threaded rods and the aluminum tie-down bars to
firmly hold FAST onto the blocks.
- Connect the control box etc.
I used the FAST control box to manually open the shutter. The
calibration lamps are too dim to be used in daylight alignment. The
laptop is needed to command the COMP mirror out of the optical path, and
to move the collimator mirror to a nominal focus position.
- Align the laser.
I used the laser from the 60-inch alignment tool. The laser
should shine through the center of the slit, reflect off the center of the
collimator mirror, reflect off the center of the grating, and hit the
center of the fold mirror. I used the pinhole mounted in the spare slit
mask holder to define the center of the slit. The old collimator mirror
has an inscribed center mark. The aluminum disk with the plastic tip can
be used to inscribe a circular mark on the collimator if needed. The
grating only needs to be centered approximately, but should be tilted so
the beam falls in the center of the fold mirror. The plastic insert with
the white dot (and white plastic screw behind it) can be placed in the
fold mirror hole to mark the center of the fold mirror.
- Put in grating surrogate.
Remove one of the gratings (I chose the 1200 line, since it is
used the least) and replace it with the grating surrogate. The center
mirror on the grating surrogate was bonded into place while the whole
assembly was lying face-down on a granite table. Thus the mirror can be
assumed to be co-planar with the aluminum grating surrogate.
- Verify optic spacings.
I used the alignment rods to verify the spacings between the
optics. The brass rods are to be used in conjunction with the inside
micrometer + delrin tip and the 24.000" aluminum rod in the Hecto
spectrograph room at the MMT. The nominal spacings (from Zemax) are
summarized on the attached page.
- Replace Collimator.
The collimator mirror is accessed from the bottom plate of FAST.
Four bolts need to be unscrewed, and the (partial) ring holding the mirror
carefully pulled out on its pins (top and bottom). The collimator mirror
is aligned by shimming the three pads. Note that two radial pads are hard
points, and one radial pad is spring loaded. It is important that the
mirror be pushed firmly against the hard points to obtain repeatable
positioning. The collimator mirror should be centered to within +-1 mm,
and aligned to < 0.2 deg side-side and < 0.5 deg up-down.
- Replace the Schmidt Corrector.
The Schmidt Corrector is basically a UBK7 window with a slight
bump in the middle of one side. The corrector should be oriented with
the bump pointing towards the fold mirror. The tolerances are very
forgiving: +-5 mm centration and +- 5 deg tip/tilt. I chose to leave
the corrector mount as-is during the August 2003 alignment, since the ~3mm
de-center I observed was within spec.
- Replace the Fold Mirror.
Tilt the grating surrogate off-axis so that the laser reflects
back to the surrogate and its position can be marked. The final alignment
is verified by centering the beam on the Fold Mirror (using the white dot
insert) and making sure the laser reflects to the center of the Camera
Mirror using the mirror insert.
The fold mirror is shimmed on three pads. I found I could simply
set the mirror in place to do a rough alignment before putting the clamps
in for the final alignment.
The fold mirror is accessed from the dewar flange. The three
pinned, spring-loaded clamps are difficult to access. The two clamps on
the left-hand side (looking at the back of the mirror) have the hard
radial pads. Special care needs to be taken with the two loose springs
that spring-load the bare metal clamp on the right-hand side; when
inserting that clamp I kept the springs from falling out by using small
pieces of tape to hold them to the clamp, and then needlenose pliers to
carefully remove the tape when I was done.
- Replace the Camera Mirror.
The Camera Mirror should be aligned so that the Grating -> Fold ->
Camera retro-reflection lines up on itself. The three clamps are easily
accessed from the main side panel, and the camera mirror is shimmed on its
three pads.
- Glyptol the bolts.
To keep the bolts from loosening, dab a touch of epoxy on them.
Same for the bolts on the grating mounts.
- Pack-up and move FAST.
Wipe down the surfaces and vacuum the inside of FAST. Blow
accumulated dust off of the optics. Reverse set-up procedure and move
FAST back to dome.
- Put dewar on and take COMPS and FLATS.
I was unable to do this in August 2003, but the final check is to
take a set of focus COMPS and some FLATS to see how everything works.
1.5m Corrector Replacement / Cleaning Procedure:
This procedure is easiest when the primary mirror is disassembled
for re-aluminization, and the corrector lens barrel is easily accessible.
However, at other times, leave the mirror covers closed, tilt to az 60d,
el 45d for
easy access, remove the stove pipe and then the 4 outer screws that hold the
barrel to the cell (on the top flange, not the recessed screws).
Then, carefully lift the barrel by the handles, making
note of the alignment white marks.
Note: The top corrector lens is a convex meniscus lens that sticks up over
0.6 inch, so it is all too easy to accidentally bump its surface. Be
careful, and wear latex gloves when working with the lenses.
Carefully screw two 6-32 threaded rods or 2" bolts into the convex
meniscus lens ring. Use the 6-32 bolts to unscrew the entire corrector
lens assembly from the barrel. The plastic retainer ring
should unscrew fairly easily. Once you have removed the ring, use the
same 6-32 bolts screwed onto the corrector lens assembly to lift it
up, out of the barrel. Reverse for re-assembly.
The corrector lens assembly consists of three stainless steel
rings. The lenses are bonded into the top two rings; the bottom ring is a
spacer. Three 4-40 bolts hold the three rings together. In addition, the
rings holding the lenses are pinned.
The new corrector lenses with the Spectrum Thin Film dielectric
AR coating (harder than sol-gel) can be cleaned as follows:
1. Blow off the surfaces (i.e. with dry N2).
2. Wash surfaces with reagent-grade isopropyl alcohol and lens tissue.
NOTE: Do not squirt and let liquid run off. Instead, DAUB very lightly
and slowly with wadded-up Kim-wipe tissues. Take your time!
Squirting isopropyl alcohol on lens 1 as it was held
vertically (as per instructions from Spectrum Thin Films) resulted in
small spots on the coating, as shown
here.
The lesson is: Do not allow water to stay on the lenses!
Picture gallery:
FAST alignment
FAST alignment (on bench I)
FAST alignment (on bench II)
FAST tolerances
FAST New Cam
FAST New Coll
FAST New Fold
FAST New Old Coll