ADC Cleaning Notes
Created 05/12/14 by EF
From Gabor:
- The first step is always to blow the loose
dust off, before starting any contact cleaning process.
-
For cleaning use a 50-50% solution of methanol and acetone, and use
soft wipes, gloves. Never use the same wipe surface twice. Fold each
wipe and keep changing its side. Use a radial motion from the center.